Mass Flow Applications

Written by: Roselius, Herb on Jul 18, 2011

Mass Flow tranducers 0 to 5V output reads 0 to 2000 SCCM.

Mass flow sensors measure gas flow in many applications. They are usually mounted inside a system that very often prevents easy direct reading of the flow if the display is integral with the mass flow sensor. A typical output is 0V for 0 flow and 5V for the maximum range of the flow sensor.

App Mass Flow 01 image link to DPV-3

A simple SPAN (gain) adjustment on a DPV-3 can set 5V to display 2000 counts on the display. However, the flow displayed is only valid for a standard gas, usually nitrogen. Other gases require a conversion factor to be applied to a reading.

For example:
The actual flow rate of a particular gas = flow rate of nitrogen X correction factor.

If the correction factor is less than one, such as 0.73 for ammonia (NH3), the DPV-3 can be simply readjusted for 2000 X 0.73 = 1460 with a 5V input.

If the correction factor is greater than one, such as 1.45 for helium (He), another input voltage must be selected such that the reading is less than 2000 (above that is an over range condition for the DPV-3). 2000 X 1.45 = 2900, therefore divide by 2, and a 2.5 V input should be adjusted to display 2900/2 = 1450.

App Mass Flow 02 image

Note: the correction factors for different gases vary for each mass flow transducer and should be requested from the manufacturer.




Leave a comment:

Your Name: *
Your Email: *
Your Web Address:
Your Comments: *

* Fields marked with the red asterisk are required.

Search
Cart Instructions
cart image

Use the 'Order here...' links located through-out the product pages. Then use the checkout button to use our secure check out system.
Terms of service

cart image

Visa mastercard American Express
Credits
© 2012 Acculex Web design by Cairn Applications